Components
Moisture Sensor (external / optional)
The moisture sensor protects downstream analyzers against humidity. This monitoring device reliably
signals a condensate breakthrough in case of failure of cooling devices or failure of condensate drains,
thus avoiding costly downtime and high repair costs. In case of an alarm, sample gas pumps can be
switched off and shut-off valves can be activated. The moisture sensor operates according to the principle
of electric conductivity. Its design ensures that any condensate in the sample gas flow gets directly to the
sensor surface by gravitational force. Even smallest amounts of liquid will trigger a fast and reliable
alarm. Control/alert served by an appropriate electronic control.
FEATURES MS
Protection against condensate break-through
Reliable detection of smallest amounts of liquid
Line break monitoring
High chemical resistance
Optional electronic control