Assembly and Installation
Remote Monitoring Connection
The device has a remote monitoring option, i.e. a floating collective fault notification.
WARNING! Overcharge!
Maximum rated load for the remote monitoring connection: 1,0A
Contact Strip (optional)
WARNING! Wrong connection!
For special models the assignment of the contact strip can be
different. Please check the exact labeling of the contact strip before connecting components.
Conntecting the contact strip:
Contact 11-13
Operational monitoring of moisture guard 2 (optional)
Contact 9-10
Connection of external moisture guard 2 (optional)
Contact 6-8
Operational monitoring of moisture guard 1 (optional).
Contact 4-5
Connection of external moisture guard 1 (optional)
Contact 1-3
Standard operational monitoring of the sample gas cooler.
Contacts 4-8 are only active if your cooler model is equipped with an electronic control unit for
moisture guards (option). Contacts 9-13 are only active if your cooler is equipped with an
additional second electronic control unit (option).