SBI521 Service Manual
67
Command: N<CR> (4eh 0dh)
Response: This command is sent to the indicator to revert to gross mode. The indicator
will not respond if the scale is in motion, positive overload or negative overload. The
indicator will also not respond if it is not in gross mode or a tare has yet to be
established.
<STX> OK<CR><LF>
Command: all others
Response: Unrecognized command
<STX>?? <CR><LF>
Table 6.4 Summary of Command and Response
Command
Response
ASCII
HEX
P<CR>
W<CR>
50 0d
57 0d
Read scale weight:
<STX><POL>xxxxx.x<L/K><G/N><STAT><CR><LF> (200S)
<STX><POL>xxxxx.x<SP><LB/KG><SP><GR/NT><CR><LF> (200F)
Z<CR>
5a 0d
<STX> OK<CR><LF>; simulate ZERO key
T<CR>
54 0d
<STX>OK<CR><LF>; simulate TARE key
C<CR>
U<CR>
43 0d
55 0d
<STX> OK<CR><LF>; simulate UNIT key
G<CR>
47 0d
<STX>OK<CR><LF>;
N<CR>
4e 0d
<STX>OK<CR><LF>;
others
x
<STX>? ?<CR><LF>