![AGRI TRONIX Brecknell SBI-521 Series Service Manual Download Page 33](http://html1.mh-extra.com/html/agri-tronix/brecknell-sbi-521-series/brecknell-sbi-521-series_service-manual_2869497033.webp)
SBI521 Service Manual
31
3.7.1 Activate the Hold Function
The CONFIG-FUNC-HOLD menu item must be set to YES.
Menu items of USER-HOLD-HLD.MOD /-AVG.TIM /-HLD.TIM /-DYN.RNG /-STB.TIM,
USER-OTHER-NLD.RNG need be set to reasonable value.
To speedup sampling of weight, set USER-HOLD-AD.H.SPD menu item to YES.
3.7.2 Access the Hold Mode
To enter the HOLD mode, press the
[HOLD]
key while in the normal weighing mode,
counting mode or percent weighing mode.
3.7.3 Hold Methods
l
Positive Peak Number HOLD mode
l
Negative Peak Number HOLD mode
l
Toggle HOLD mode
l
Average HOLD mode
l
Auto HOLD mode
The following are details of these HOLD modes:
Positive Peak HOLD
When USER-HOLD-HLD.MOD is set to PS.PEAK, the hold mode is positive peak hold
mode.
When scale first enters this working mode, it will display the largest positive number
that is from the time of zero-point set. After entering this working mode, scale will
always catches and refresh positive larger number and display it.
To exit HOLD mode, press the
[HOLD]
key again.
Negative Peak HOLD
When USER-HOLD-HLD.MOD is set to NG.PEAK, the hold mode is negative peak hold
mode.
When scale first enters this working mode, it will display the largest negative number
from the time of zero-point set. After entering this working mode, scale will always
catches negative larger number and display it.
To exit HOLD mode, press the
[HOLD]
key again.
Toggle HOLD
When USER-HOLD-HLD.MOD is set to TOGGLE, the hold mode is toggled between
HOLD and MANUAL HOLD function.
After entering this working mode, the scale will freeze and display number if scale is
stable. Only the weight that is over USER-OTHER-NLD.RNG (zero 'dead' band) can
be held.