Operational Check
Operational Check
This section provides operational check to confirm the test set and analyzer operational
performance. The operation verification limits provided ensure that your test set and
analyzer are operating properly.
Verification Limits
Specifications for the test set are typical. System performance for the analyzer and test set
are only characteristic and intended as non warranted information. Only a functional
certificate is provided for the U3042AE08.
It is recommended that you return your instrument to Agilent Technologies for servicing or
repair if the test set and analyzer performance exceed the operational verification limits.
NOTE
Typical specifications are based on 1 to 2 units performance. Refer to
Table 14
and
Table 15 on page 46
.
Table 14
N5230C Operation Verification Limit for Reflection Tracking
1
1.
Reflection Tracking takes into account Source Loss, Receiver
Loss, Margin, and analyzer’s Mixer Cal.
Frequency
Standard
Port 1-12
2
2. A standard unit is one without Options 001 or 002.
Option 001 or 002
Port 1-12
10 MHz to 4 GHz
−
10.0 dB
0 dB
4 GHz to 6 GHz
−
13.0 dB
−
2.5 dB
6 GHz to 10 GHz
−
16.0 dB
−
5.0 dB
10 GHz to 15 GHz
−
17.0 dB
−
7
.
5 dB
15 GHz to 20 GHz
−
25.0 dB
−
15.0 dB