User’s Guide U3042-90002
49
U3042AE08
Operational Check
Figure 43
Calibration Complete
6. Repeat
step 1
thru
step 5
(1-Port Calibration Procedure) for Ports 2 thru 12. When
finished, there should be sixteen Cal Sets saved with the titles “999.1” thru “999.12”
(12-Port).
If you are using an ECal module you can verify the individual port calibration by
selecting
Response
>
CAL
>
More
>
ECAL
>
ECAL Confidence Check
. For further
information refer to the Help menu.
7. Press
Trace/Chan
>
Trace
>
Delete Trace
. There should be no traces on the display.