User’s Guide U3042-90002
5
U3042AE08
General Specifications
General Specifications
Specifications for the test set are characteristic for the system performance of the analyzer
and test set. Actual performance of the system is based on the customers analyzer and
options that are used with the test set. A functional certificate is only offered for the
U3042AE08.
NOTE
When connected to a network analyzer, this test set will degrade the
performance at the test ports. The internal switch paths reduce test port
power and power to the receivers. This affects the test port power of the
analyzer and also reduces dynamic range. The reflection tracking values
measured in the
“Operational Check” on page 45
can be subtracted from the
analyzers dynamic range to determine the approximate performance of the
system.
Power Requirements
Verify that the required ac power is available before installing the test set to the analyzer.
• 100/120/220/240 VAC (50/60Hz)
• The instruments can operate with mains supply voltage fluctuations up to ± 10% of
the nominal voltage.
• Air conditioning equipment (or other motor–operated equipment) should not be placed
on the same ac line that powers the test set and analyzer.
•
Table 1
contains the maximum wattage for all instruments. This table can be use to
determine the electrical and cooling requirements.
WARNING
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall be
inserted only into a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the instrument, is likely to make the instrument dangerous.
Intentional interruption is prohibited.
Table 1
Power Requirements
Standard Equipment
Instrument
Maximum
Wattage
N5242A
450
N5230C
350
U3042AE04
350