18
User’s Guide U3042-90002
U3042AE08
System Setup with N5242A
6. Secure the network analyzer’s lower locking feet to the test set upper locking feet, using
the spring–loaded screws on the locking feet. Refer to
Figure 16
. If the network
analyzer's lower locking feet are not aligned with the screw holes in the test set's upper
locking feet, loosen the screws securing the feet to the instrument slightly to align.
Figure 16 Locking Feet Screws
NOTE
There are two Lock-Feet kits available. Refer to
“Agilent Support, Services,
and Assistance” on page 68
for ordering information.
• Analyzer
−
5023-0132 (Kit includes locking feet and screws)
• Test Set
−
N5242-20138 is the right foot and the N5242-20139 is the left
foot.
• Screw
−
0515-2317
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html