Calibrator Output Operations
2
U1401A User’s and Service Guide
31
Memory Generation
For constant voltage and current outputs, the U1401A offers
two additional useful functions. One is an
autoscan
output
that is able to generate up to 16 different steps of constant
voltage or current each with its own user- defined amplitude
and time interval. The other one is an
autoramp
output
with user- defined dual slopes and number of steps for linear
simulation.
Autoscan output
To set the
autoscan
output:
1
Turn the rotary switch to any one of the
(constant
current output) or
(constant voltage output)
positions.
2
Press
SHIFT
to access the shifted operations of the keypad.
The
annunciator will appear on the display.
3
Follow one of the instructions below:
•
For voltage output, press
MODE
to cycle through ±1.5 V,
±15 V,
±1.5 V,
±15 V,
±1.5 V, and
±15 V output modes. Select one of the two
output modes, depending on the voltage range
you require.
•
For current output, press
MODE
to cycle through
±25 mA,
±25 mA, and
±25 mA output
modes. Select the
output mode.
Summary of Contents for U1401A
Page 1: ...Agilent Technologies U1401A Handheld Multi Function Calibrator Meter User s and Service Guide ...
Page 16: ...XVI U1401A User s and Service Guide Contents ...
Page 46: ...26 U1401A User s and Service Guide 1 Getting Started ...
Page 64: ...44 U1401A User s and Service Guide 2 Calibrator Output Operations ...
Page 67: ...Making Measurements 3 U1401A User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Page 104: ...84 U1401A User s and Service Guide 4 Changing the Default Settings ...
Page 115: ...Application Examples 5 U1401A User s and Service Guide 95 Figure 5 5 Zener diode test ...
Page 146: ...126 U1401A User s and Service Guide 6 Maintenance ...