30
Chapter 2
Test Descriptions
Input VSWR
Step 9.
Select the relevant calibration kit in use from the Cal menu, as the calibration kit
selection is dependant on the NFA being tested (press the
Cal
key,
Calibration Kit
selection and
Return
).
NOTE
The connector type applies to the test port or adaptor if fitted and not the NFA under
test.
Step 10.
Select an S11 port calibration (press the
Cal
key, then
Calibrate Menu
and
Reflection 1-port
).
Step 11.
Connect the Open to the test port or, if fitted, the adapter and perform the open
calibration (press
Opens
and then select the relevant menu key).
Step 12.
Connect the Short to the test port or, if fitted, the adapter and perform the open
calibration (press
Shorts
and then select the relevant menu key).
Step 13.
Connect the Load to the test port or, if fitted, the adapter and perform the open
calibration (press
Load
, select the relevant menu key
and then
done 1-port CAL
).
Step 14.
Save the calibration (press the
Save/recall
key and then
Save State
).