Chapter 2
61
Test Descriptions
Noise Figure Range and Accuracy
Equation 1
Range
dB
0.125
= Metrology Data
dB
at 1dB / 8
Range
dB
0.5
= Metrology Data
dB
at 1dB / 2
Range
dB
1 to 11 = Metrology Data
dB
from 1 to 11
Range
dB
12 to 22 = Metrology Data
dB
from 1 to 11 + (11dB to re-reference)
Equation 2
Range
dB
0.125
= Measured Attenuation
dB
at 1dB / 8
Range
dB
0.5
= Measured Attenuation
dB
at 1dB / 2
Range
dB
1 to 11 = Measured Phot
dB
Ref 1 - Measured Phot
dB
from -55 to -65
Range
dB
12 to 22 = Measured Phot
dB
Ref 2 - Measured Phot
dB
from -66 to -76 +
(11dB to re-reference)
NOTE
The ENR values given in the calculations are taken to be 5dB, 15dB and 22dB
respectively. The actual and measured attenuations supplied within the appropriate
Test Records are calculated from in the equations in the Tables on page 62.
8
-73
19
9
-74
20
10
-75
21
11
-76
22
Table 2-13
Noise Figure Range and Accuracy Test Worksheet
Attenuator
Step Size (dB)
Input
Level
(dBm)
Measured
Phot (dB)
Range
(dB)
Actual
Attenuation (dB)
Equation 1
Measured
Attenuation (dB)
Equation 2
Summary of Contents for NFA Series
Page 114: ...104 Chapter3 Technical Specifications General Specifications...
Page 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Page 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Page 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Page 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Page 184: ...174 Index Index...