Chapter 2
67
Test Descriptions
Gain Measurement Uncertainty
Step 9.
Convert the delta ratios to dB error values using 10*LOG
10
(Delta reading from
index 1 to 71). Enter the value in the Worksheet No. 1.
Example: Index 16 = 1.004545 = 10 * LOG
10
(1.004545) = 0.019694dB
Step 10.
Using the error data calculated in the Worksheet No. 1, enter the peak-to-peak error
in dB over the 30 available measurement ranges in the Worksheet 2.
Step 11.
Record the worst Instrumentation Uncertainty into the Test Record and ensure that
the measured value is within its published specification.
Summary of Contents for NFA Series
Page 114: ...104 Chapter3 Technical Specifications General Specifications...
Page 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Page 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Page 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Page 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Page 184: ...174 Index Index...