![Agilent Technologies NFA Series Performance Verification And Calibration Manual Download Page 38](http://html1.mh-extra.com/html/agilent-technologies/nfa-series/nfa-series_performance-verification-and-calibration-manual_2868187038.webp)
26
Chapter 2
Test Descriptions
Input VSWR
VSWR Test Specification
Related Adjustments
None
Required Test Equipment
N8973A
N8974A
N8975A
10 MHz to 500 MHz
< 1.5:1
< 1.5:1
< 1.5:1
> 500 MHz to 1500 MHz
< 1.7:1
< 1.7:1
< 1.7:1
> 1500 MHz to 3000 MHz
< 1.8:1
< 1.8:1
< 1.8:1
> 3000 MHz to 6700 MHz
N/A
< 1.3:1
< 1.3:1
> 6700 MHz to 20000 MHz
N/A
N/A
< 2.1:1
> 20000 MHz to 26500 MHz
N/A
N/A
< 2.4:1
Table 2-6
Test Equipment required for Input VSWR test
Equipment description
Critical specification for
equipment substitution
Recommended model
Vector Network Analyzer 1
10 MHz to 3 GHz
8753ES or 8753ET, Option 004
a
Vector Network Analyzer 2
3 GHz to 26.5 GHz
8722ES or 8722ET, Option 004
b
N Type Calibration Kit
Impedance: 50
Ω
Frequency Band: 10 MHz to
3 GHz
c
UUT: N8973A
85032B
3.5mm Calibration Kit
Impedance: 50
Ω
Frequency Band: 10 MHz to
3 GHz
UUT: N8974A and N8975A
85033D
Summary of Contents for NFA Series
Page 114: ...104 Chapter3 Technical Specifications General Specifications...
Page 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Page 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Page 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Page 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Page 184: ...174 Index Index...