5
Key Reference
134
N9342C/43C/44C User’s Guide
Cable & Antenna Test
Key access:
[MODE] > {Cable&Antenna Test}
Select this mode to make a measurement using the
cable and antenna test function.
Auto Scale
Key access:
[Amptd] > {Auto Scale}
Press this softkey to adjust the amplitude scale
automatically to include the peak and valley level
in graticule.
Calibrate
Key access:
[Freq] > {Calibrate}
Cal Type
Key access:
[Freq] > {Cal Type}
This softkey toggles the calibration type between
full span calibration and selected span calibration.
Full span calibration runs an open- short- load
calibration across the full frequency span.
A FullSpan calibration ignores the frequency
changes during measurements.
Selected span calibration performs an
open- short- load calibration across a user- defined
frequency span. A SelectedSpan Cal requires a
re- calibration of the analyzer, after changing the
measurement frequency range. Normally, the
selected span should be set as close to the actual
swept frequency range.
Run/Hold
Key access:
[Sweep] > {Run/Hold}
This softkey toggles the frequency sweep between
run and pause status.