Key Reference
5
N9342C/43C/44C User’s Guide
129
•
Zero Type
Toggles the zeroing type of the Power Sensor
between internal and external. When this type is
set to
EXT
, pressing
Zeroing
will bring up a
message which indicates the external signal
need be pulled out. This setting is not affected
by the preset in Power Meter menu.
Meas Disp
Key access:
[MODE] > {Power Meter}
Access the Meas Disp submenu for defining the
display range and resolution while in Power Meter
mode.
• Disp Range
Access the associated Disp Range submenu.
•
Auto Range
Toggles the auto range function between On and
Off. When the
Auto Range
is set to On, the Top
and Bottom value of the power meter’s display
range will be set automatically according to the
measurement data:
Top = Max. data + 0.5
Bottom = Min. data - 0.5
The default setting is Off.
•
Top
Set the top value of the power meter’ display
range.
Range from ( 0.01) to 200 dBm.
The default value is 20 dBm.
•
Bottom
Set the bottom value of the power meter’ display
range.
Range from - 200 to (Top - 0.01) dBm.
The default value is - 70 dBm.