5
Error Messages
146
N9342C/43C/44C User’s Guide
–144 Character data too
long
Indicates the character data contained
more than 12 characters. (see IEEE
488.2, 7.7.1.4)
764
Unable to save
file
Indicates a failure occurred while saving
a file. The file was not saved.
762
Unable to load
file
Indicates a failure occurred while load-
ing a file. The file was not loaded.
612
File NOT found
Indicates the analyzer could not find the
specified file.
173
Lame package,
please upgrade
firmware
Indicates the current firmware is not
mostly updated. The firmware needs
upgrading.
172
Option install
fail, invalid
option licence
Indicates the option could not be
installed, because of the invalid option
licence.
171
File loaded
Indicates the file loading succeeded.
170
The menu is
unavailable in
this case
Indicates the memo is invalid in this
case.
166
Marker counter
opened in fast
sweep
Indicates the counter’s accuracy
decreases when in fast sweep mode.
162
Cannot open fast
sweep in this case
Indicates current settings do not allow
you to open fast sweep. For example,
currently in FFT, zero span, or any mea-
surement in power suite does not allow
fast sweep.
156
Incorrect
alignment file on
flash
Indicates an invalid alignment file.
153
RF EEPROM operate
fail
Indicates an error when writing align-
ment file.
152
Alignment file
oversize
Indicates the alignment file has error or
invalid data.
151
DSP boot fail
Indicates the analyzer cannot process
measurement currently, wait the
analyzer to reboot. If analyzer froze, try
to restart the analyzer.
Error
Code
Error Message
Description
Summary of Contents for N9342C
Page 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Page 8: ......
Page 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Page 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Page 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...