3
Functions and Measurements
60
N9342C/43C/44C User’s Guide
Setting the Power Meter’s Upper and Lower Limits
The internal Power Meter can be configured to
detect when a measurement has failed a user
predefined upper and lower limits.
1
Press
[Preset]
to perform a factory preset.
2
Press
[MODE]
>
{Power Meter} > [ENTER]
to turn on
the power meter.
3
Press
{Meas Setup}
>
{Limits}
to access the limits
menu.
4
Press
{Limits}
to activate the limits function. Each
time the softkey is pressed, the selected option
changes.
5
Press
{Upper Limits}
and enter the high limit value
using the numeric keypad. Then press
{dBm}
to
complete the setup.
Press
{Lower Limits}
and enter the low limit value
using the numeric keypad. Then press
{dBm}
to
complete the setup.