Service Guide N5230-90014
4-27
PNA Series Microwave Network Analyzers
Troubleshooting
N5230A
Measurement System Troubleshooting
EEPROM Headers
The network analyzer application uses the firmware revision information stored in the pc
board header EEPROM. If the information stored in any EEPROM is incorrect, the
network analyzer may not operate properly.
The following table lists the pc boards in your network analyzer that contain EEPROM
headers. The pc boards are listed by name and part number and the correct firmware
revision code is given for each.
To view this EEPROM header information on the network analyzer display, select
System
,
Service
,
Utilities
,
View
EEPROM
Headers
in the network analyzer application.
If the information is incorrect for any of the pc boards, contact Agilent Technologies. Refer
to
“Contacting Agilent” on page 2-11
.
Error Messages
SOURCE UNLEVELED:
The source ALC circuit on the A16 test set motherboard is running
open-loop. Check the cable connections for W30, W31, and W32 between the A19 SSLAM
and the A16 test set motherboard. Refer to those sections of this chapter that pertain to
troubleshooting the source section of the analyzer.
PC Board Name
PC Board Part
Number
Memory
ID
Hardware
ID
Firmware
Revision
A5 SPAM
E8364-63187
3
5
A
A6 and A8 multiplier
E8364-63182
3
6
A
A7 and A9 fractional-N synthesizer
E8364-63183
3
7
C
A10 frequency reference
E8364-60136
3
10
B
A16 test set motherboard
N5230-63078
3
16
D
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...