Service Guide N5230-90014
3-25
PNA Series Microwave Network Analyzers
Tests and Adjustments
N5230A
System Verification
If the System Fails the Verification Test
IMPORTANT
Inspect all connections.
Do not
remove the cable from the analyzer test port.
This
will invalidate
the calibration that you performed earlier.
1. Disconnect and clean the device that failed the verification test.
2. Reconnect the device making sure that all connections are torqued to the proper
specifications.
3. Measure the device again.
4. If the analyzer still fails the test, check the measurement calibration by viewing the
error terms as described in
in
5. Refer to
for additional troubleshooting steps.
Figure 3-10 System Verification Failure Flowchart