Service Guide N5230-90014
7-39
PNA Series Microwave Network Analyzers
Repair and Replacement Procedures
N5230A
Removing and Replacing the A41 Hard Disk Drive (HDD)
3. Point to
Programs
,
Accessories
, and then open
Windows
Explorer
.
4. Navigate to
C:\Program Files\Agilent\Network Analyzer
.
5. Copy each of the backup files from the floppy disk or USB memory drive to the HDD.
6. Navigate to
D:\CalFiles\.
7. Copy the backup files prefixed with
mxcalfile_
from the floppy disk or USB memory
drive to the HDD.
8. After all files have been copied, remove the storage device containing the backup files:
either the floppy disk from the analyzer’s disk drive, or the USB memory drive from the
USB port.
9. Exit
Windows Explorer
.
Post-Repair Procedures
1. Adhere the new Certificate of Authenticity (license) label to the outer cover in the
location specified in
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
.
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...