Service Guide N5230-90014
3-35
PNA Series Microwave Network Analyzers
Tests and Adjustments
N5230A
Adjustments
Adjustments
These adjustments are firmware-driven tests that are used to fine-tune your analyzer.
If multiple adjustments are to be performed, perform them in the order listed.
•
“10 MHz Frequency Reference Adjustment” on page 3-35
•
“LO Power Adjustment” on page 3-36
•
“Source Calibration Adjustment” on page 3-37
•
“Receiver Calibration Adjustment” on page 3-38
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network
analyzer’s 10 MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the Frequency Adjustment at 10 MHz
Procedure
NOTE
This adjustment typically adjusts to within
±
0.01 ppm.
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the frequency counter.
Figure 3-13 Setup for Adjustment of the 10 MHz Frequency Reference
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
10 MHz Freq Adjust
.
3. Ensure the GPIB settings are correct.
4. Click
Begin Adj
, and then follow the instructions as they are displayed.
Equipment Type
Model or Part Number
Alternate Model or Part Number
Cable, BNC, 50
Ω
, 24 inch
8120-1839
Any
Frequency counter
53151A, Option 001
Any that will measure a signal at 10 MHz.
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...