Service Guide N5230-90014
4-39
PNA Series Microwave Network Analyzers
Troubleshooting
N5230A
Measurement System Troubleshooting
2. If the signal is not present, replace the A7 fractional-N synthesizer board. Refer to
“Removing and Replacing the A5 through A10 Boards” on page 7-16
.
3. If the signal is present, reconnect cable W1, and then continue with
Multiplier Output, All Bands” on page 4-40
.
Table 4-7
LO Path Frequencies
Band
Network Analyzer
Frequency Band of
Problem (GHz)
Network Analyzer
Source Frequency
Setting (GHz)
Expected A7 Output
Frequency (GHz)
a
a. With the analyzer’s spur avoidance function OFF.
Expected A6 Output
Frequency (GHz)
0
.0003 to 0.001
0.0007
0.0084
0.0084
1
0.001 to 0.010
0.006
0.014
0.014
2
0.010 to 0.040
0.030
0.038
0.038
3
0.040 to 0.748
0.400
0.408
0.408
4
0.748 to 1.500
1.200
1.208
1.208
5
1.500 to 3.125
2.300
2.308
2.308
6
3.125 to 4.167
3.600
1.804
3.608
7
4.167 to 5.250
4.800
2.404
4.808
8
5.250 to 6.250
5.800
2.904
5.808
9
6.250 to 8.333
7.300
1.827
7.308
10
8.333 to 10.500
9.500
2.377
9.508
11
10.500 to 12.500
11.500
1.918
3.836
12
12.500 to 13.500
13.000
2.168
4.336
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...