
Service Guide N5230-90014
3-39
PNA Series Microwave Network Analyzers
Tests and Adjustments
N5230A
Adjustments
Figure 3-15 Setup 1 for the Receiver Calibration Adjustment
2. On the
System
menu, point to
Service
,
Adjustments
, and then click
Receiver Calibration
.
3. Ensure the GPIB settings are correct.
4. Click
Calibrate
, and then follow the instructions as they are displayed.
Figure 3-16 Setup 2 for the Receiver Calibration Adjustment
Summary of Contents for N5230A
Page 10: ...Contents 6 Service Guide N5230 90014 Contents...
Page 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Page 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Page 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Page 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Page 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Page 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Page 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Page 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...