MEASure
The
MEASure
command subsystem is an HP E1326B/E1411B Multimeter
subsystem which configures the multimeter for strain measurements. After
configuring the multimeter,
MEASure
immediately makes the (strain)
measurement.
Executing
MEASure
is equivalent to configuring the multimeter with the
low-level commands shown in the following table.
Parameter
Command
Setting
Range
VOLTage:RANGe
RESistance:RANGe
As specified or autorange.
Resolution
VOLTage:RESolution
RESistance:RESolution
As specified, or as a function of range,
integration time, or aperture time.
Aperture Time
VOLTage:APERture
RESistance:APERture
16.7 ms (60 Hz) or 20 ms (50 Hz), or based on
specified resolution.
Integration Time
VOLTage:NPLC
RESistance:NPLC
1 power line cycle (PLC) or based on specified
resolution.
Autozero
CALibration:ZERO:AUTO
ON (autozero is performed after every
measurement).
Offset Compensation
RESistance:OCOMpensated
OFF (resistance measurements only).
Trigger Source
TRIGger:SOURce
IMM (trigger signal is always true).
Measurement is taken when multimeter goes
to trigger state.
Number of Triggers
or Number of Scans
TRIGger:COUNt
1; number of triggers issued or number of
scans through
channel list
before multimeter
returns to idle state.
Trigger Delay
TRIGger:DELay
AUTO (delays are 0 seconds for DC voltage
and resistance; 0.5 seconds for AC voltage).
Readings per Trigger
SAMPle:COUNt
1; number of measurements taken when
trigger is received.
Sample Period
SAMPle:SOURce
IMM; period between measurements or the
period between FET multiplexer scans.
Refer to Chapter 5 in the HP E1326B/E1411B Multimeter User’s Manual
for more information on the
MEASure
subsystem.
Subsystem Syntax
MEASure
:STRain [:<
strain_function
>]? <
channel_list
>
Chapter 5
Strain Gage Multiplexer Command Reference 77