Closing Channels - FET multiplexers: To close:
–
a single channel, use
CLOSe
(
@ccnn
);
–
single channels on different cards, use
CLOSe
(
@ccnn,ccnn
).
Closure order for multiple channels with a single command is not
guaranteed.
•
Related Commands:
[ROUTe:]OPEN
,
[ROUTe:]CLOSe?
,
[ROUTe:]SCAN:MODE
,
[ROUTe:]SCAN:PORT
•
*RST Condition: All multiplexer channels are open.
Example
Closing Multiplexer Channels
These examples close channel 00 of switchbox multiplexer card number 1.
Relay Multiplexer Switchbox:
CLOS (@100)
Closes channel 0 of multiplexer #1.
FET Multiplexer Switchbox:
SCAN:MODE VOLT
Selects the voltage scanning mode.
SCAN:PORT ABUS
Selects the analog bus port.
CLOS (@100)
Closes channel 0 of multiplexer #1.
CLOSe?
[ROUTe:]CLOSe?
<
channel_list
>
returns the current state of the channel
queried. The channel_list is in the form (
@ccnn
). The command returns 1
if the channel is closed or returns 0 if the channel is open.
Comments
•
Query is Software Readback: The
[ROUTe:]CLOSe?
command
returns the current software state of the channel specified. It does
not account for relay hardware failures.
Example
Query Multiplexer Channel Closure
CLOS (@100)
100 closes channel 0 of multiplexer
#1.
CLOS? (@100)
Query channel 0.
Chapter 5
Strain Gage Multiplexer Command Reference 81