:CONTinuous?
INITiate:CONTinuous?
queries the scanning state. With continuous
scanning enabled, the command returns 1. With continuous scanning
disabled, the command returns 0.
Example
Query Continuous Scanning State
INIT:CONT ON
Enables continuous scanning.
INIT:CONT?
Query continuous scanning state.
[:IMMediate]
INITiate[:IMMediate]
starts the scanning cycle and closes the first channel
in the channel list. Successive triggers from the source specified by the
TRIGger:SOURce
command advance the scan through the channel list.
Comments
•
Starting the Scanning Cycle: The
INITiate[:IMMediate]
command
starts scanning by closing the first channel in the channel list. A
trigger advances the scan through the channel list. An invalid
channel list generates an error (see the
[ROUTe:]SCAN
command).
•
Stopping Scanning Cycles: See the
ABORt
command.
Example
Enabling a Single Scan
SCAN (@100:107)
Sets channel list.
INIT
Starts scanning cycle.
76 Strain Gage Multiplexer Command Reference
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com