:STRain
CONFigure:STRain[:
<
strain_function
>
]
<
channel_list
>
configures the
multimeter for strain measurements using the specified bridge arrangement
on the specified channel.
Parameters
Parameter Name
Parameter Type
Range of Values
strain_function
Discrete
[QUARter]
HBENding
HPOisson
FBENding
FPOisson
FBPoisson
QTENsion
QCOMpression
UNSTrained
channel_list
Numeric
00 - 07
Comments
•
The strain measurements specified by the strain_function parameter
are:
QUARter
=
1
⁄
4
bridge measurements
HBENding
= Bending
1
⁄
2
bridge measurements
HPOisson
= Poisson
1
⁄
2
bridge measurements
FBENding
= Bending Full bridge measurements
FPOisson
= Poisson Full bridge measurements
FBPoisson
= Bending Poisson Full bridge measurements
QTENsion
= Tension shunt diagnostic
QCOMpression
= Compression shunt diagnostic
UNSTrained
= Unstrain reference voltage measurement
•
The default strain_function is
QUARter
(
1
⁄
4
bridge measurements).
•
QTENsion
and
QCOMpression
are used for relay strain gage
multiplexers only.
Example
Making
1
⁄
4
Bridge Strain Measurements
STR:GFAC 2.11E-6,(@100:103)
Specify the gage factors for the
channels.
CAL:STR (@100:103)
Measure the unstrained references.
CONF:STR:QUAR (@100:103)
Configure the multimeter and make
1
⁄
4
bridge measurements.
VOLT:RANG MIN
Set 0.125V range.
72 Strain Gage Multiplexer Command Reference
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com