Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
89
Optimizing the Detector
5
Getting Started and Checkout
Setting the Chromatographic Conditions
1
Set up the system with the following chromatographic conditions and wait
until the baseline gets stable.
Table 6
Chromatographic Conditions
Mobile phases
A = water = 35 %
B = Acetonitrile = 65 %
Column
OSD-Hypersil column, 125 mm x 4 mm i.d. with 5
µm particles
Sample
Isocratic standard sample, 1:10 diluted in
methanol
Flow rate
1.5 ml/min
Compressibility A (water)
46
Compressibility B (Acetonitrile)
115
Stroke A and B
auto
Stop time
4 min
Injection volume
5 µl
Oven temperature (1200)
30 °C
FLD Excitations/Emission Wavelength
EX = 246 nm, EM = 317 nm
FLD PMT Gain
PMT = 10
FLD Response time
4 s