![Agilent Technologies G7121A User Manual Download Page 82](http://html.mh-extra.com/html/agilent-technologies/g7121a/g7121a_user-manual_2867752082.webp)
82
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
4
Preparing the Module
Solvent Information
PEEK
PEEK (Polyether-Ether Ketones) combines excellent properties regarding
biocompatibility, chemical resistance, mechanical and thermal stability. PEEK
is therefore the material of choice for UHPLC and biochemical
instrumentation.
It is stable in the specified pH range (for the Bio-inert LC system: pH 1 – 13,
see bio-inert module manuals for details), and inert to many common solvents.
There is still a number of known incompatibilities with chemicals such as
chloroform, methylene chloride, THF, DMSO, strong acids (nitric acid > 10 %,
sulphuric acid > 10 %, sulfonic acids, trichloroacetic acid), halogenes or
aequous halogene solutions, phenol and derivatives (cresols, salicylic acid
etc.).
When used above room temperature, PEEK is sensitive to bases and various
organic solvents, which can cause it to swell. Under such conditions normal
PEEK capillaries are very sensitive to high pressure. Therefore Agilent uses
stainless-steel cladded PEEK capillaries in bio-inert systems. The use of
stainless steel cladded PEEK capillaries keeps the flow path free of steel and
ensures pressure stability to at least 600 bar. If in doubt, consult the available
literature about the chemical compatibility of PEEK.
Polyimide
Agilent uses semi-crystalline polyimide for rotor seals in valves and needle
seats in autosamplers. One supplier of polyimide is DuPont, which brands
polyimide as Vespel, which is also used by Agilent.
Polyimide is stable in a pH range between 1 and 10 and in most organic
solvents. It is incompatible with concentrated mineral acids (e.g. sulphuric
acid), glacial acetic acid, DMSO and THF. It is also degraded by nucleophilic
substances like ammonia (e.g. ammonium salts in basic conditions) or
acetates.
Polyethylene (PE)
Agilent uses UHMW (ultra-high molecular weight)-PE/PTFE blends for yellow
piston and wash seals, which are used in 1290 Infinity pumps, 1290 Infinity II
pumps, the G7104C and for normal phase applications in 1260 Infinity pumps.
Polyethylene has a good stability for most common inorganic solvents including
acids and bases in a pH range of 1 to 12.5. It is compatible to many organic