Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual
161
Error Information
7
Detector Error Messages
Motor Errors
Motor Error
Error ID: EX 6700, EM 6701
Motor Or Encoder Not Found
Error ID: EX 6705, EM 6706
During initialization of the detector, the excitation and emission
monochromator are activated.
N O T E
Monochromator motor errors may show up duringthe
initialization
or during
operation
ofthe detector.There are individual messages for either the excitation or the
emissionside.If an error occurs, do a lamp ignition. This will clear the errorand a
re-initialization of the motors is performed.
Probable cause
Suggested actions
1
Friction too high.
Exchange the monochromator assembly.
2
Defective monochromator assembly.
Exchange the monochromator assembly.
Probable cause
Suggested actions
1
Encoder cables mixed on FLM board.
Check encoder connections to FLM.
2
Monochromator assembly not connected.
Check motor connections to FLF (Ex) and FLM
(Em) and encoders to FLM.
3
Monochromator or encoder defective.
Replace monochromator assembly.
4
Monochromator motor power driver
defective.
Defective FLF board (Ex) or FLM board (Em).