Early Maintenance Feedback
Early Maintenance Feedback
Maintenance requires the exchange of components which are subject to wear
or stress. Ideally, the frequency at which components are exchanged should be
based on the intensity of usage of the module and the analytical conditions,
and not on a predefined time interval. The early maintenance feedback (
EMF
)
feature monitors the usage of specific components in the instrument, and
provides feedback when the user-selectable limits have been exceeded. The
visual feedback in the user interface provides an indication that maintenance
procedures should be scheduled.
EMF Counters
EMF counters
increment with use and can be assigned a maximum limit which
provides visual feedback in the user interface when the limit is exceeded.
Some counters can be reset to zero after the required maintenance procedure.
Using the
EMF Counters
The user-settable
EMF
limits for the
EMF Counters
enable the early maintenance
feedback to be adapted to specific user requirements. The useful maintenance
cycle is dependent on the requirements for use. Therefore, the definition of the
maximum limits need to be determined based on the specific operating
conditions of the instrument.
Setting the EMF Limits
The setting of the
EMF
limits must be optimized over one or two maintenance
cycles. Initially the default
EMF
limits should be set. When instrument
performance indicates maintenance is necessary, take note of the values
displayed by the
EMF counters
. Enter these values (or values slightly less than
the displayed values) as
EMF
limits, and then reset the
EMF counters
to zero.
The next time the
EMF counters
exceed the new
EMF
limits, the
EMF
flag will be
displayed, providing a reminder that maintenance needs to be scheduled.
The detector provides the following EMF counters:
•
Flash Lamp Life Time
N O T E
This function is only available via Agilent Lab Advisor or Instant Pilot.