Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
136
9
Maintenance
Replace Valve Heads
CAUTION
Column Damage or Bias Measurement Results
Switching the valve to a wrong position can damage the column or bias
measurement results.
Fit the lobe to the groove to make sure the valve is switched to the correct
position.
CAUTION
The valve actuator contains sensitive optical parts, which need to be protected
from dust and other pollution. Pollution of these parts can impair the accurate
selection of valve ports and therefore bias measurement results.
Always install a valve head for operation and storage. For protecting the
actuator, a dummy valve head (part of Transportation Lock Kit
(G1316-67001 )) can be used instead of a functional valve. Do not touch
parts inside the actuator.
NOTE
The tag reader reads the valve head properties from the valve head RFID tag
during initialization of the module. Valve properties will not be updated, if the
valve head is replaced while the module is on. Selection of valve port positions
can fail, if the instrument does not know the properties of the installed valve.
NOTE
To have the valve correctly recognized by the Agilent Infinity Valve Drive you
must have the valve drive powered off for at least 10 seconds.