Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
77
3
Using the Module
Installing and Using Column ID Tags
Figure 16 Column Assignment dialog box overview
The
Column Assignment
dialog box has three sections that give you
information about the column:
•
The
Plumbing
section shows the valve connections to the positions in the
column compartment.
•
The
Visualization
section gives a visual representation of the configuration
of the columns in the column compartment; the columns are color coded.
•
Place the mouse cursor over a column to display a tooltip of the
column information from the column ID tag.
•
Click a column to highlight the column information in the
Column Tag
Information
table.
•
The
Column Tag Information
shows the information in the column ID tags
for all columns in the configuration, including their location in the column
compartment and their color codes.
NOTE
Make the connections to give the shortest distances between the valve ports and
the columns, and use a logical order (left column 1 to port 1-1', left column 2 to
port 2-2' and so on). Avoid leaving unused ports between used ones.