Agilent InfinityLab LC Series Multicolumn Thermostats User Manual
81
3
Using the Module
Installing and Using Column ID Tags
Column Tag Information
The column ID tag contains the following information:
Field
Description
Type
Read/Write
permission
pre-labeled tag
Read/Write
permission
custom tag
Description
A description of the column.
Static
Read
Write
Length
The length of the column in mm.
Static
Read
Write
Diameter
The internal diameter of the column in mm.
Static
Read
Write
Particle Size
The particle size of the column packing mate
-
rial in µm.
Static
Read
Write
Maximum Pressure
The maximum pressure supported by the col
-
umn.
Static
Read
Write
Number of Injections
The number of injections that have been made
on the column.
Dynamic
Read
Read
Product Number
The product number of the column.
Static
Read
Write
Maximum Measured
Temperature
The highest temperature (in °C) experienced by
the column to date.
Dynamic
Read
Read
Maximum Tempera
-
ture
The safe maximum operating temperature of
the column (in °C).
Static
Read
Write
Minimum pH
The minimum pH supported by the column.
Static
Read
Write
Maximum pH
The maximum pH supported by the column.
Static
Read
Write
Void Volume (mL)
The void volume of the column and fittings.
Static
Write
Write
First Injection
The date and time of the first injection onto the
column.
Dynamic
Read
Read
Recent Injection
The date and time of the most recent injection
onto the column.
Dynamic
Read
Read
Manufacturing Date
The date of manufacture of the column.
Static
Read
Write
Agilent Column
Whether or not the column was supplied by
Agilent Technologies.
Static
Read
Write
Serial Number
The serial number of the column.
Static
Read
Write
Batch Number
The batch number of the column.
Static
Read
Write
Tag Sealed
Whether or not all static fields except
Comment
and
Void Volume
are set irrevocably to
read-only.
Static
Read
Write
Comment
A user-generated comment about the column. Static
Write
Write