Agilent 1260 Infinity VWD User Manual
95
Error Information
7
Detector Error Messages
Ignition Failed
Error ID: 7452
The lamp failed to ignite. The processor monitors the lamp current during
the ignition cycle. If the lamp current does not rise above the lower limit
within 2 – 5 s, the error message is generated.
Probable cause
Suggested actions
1
Defective or non-Agilent lamp.
Exchange the lamp.
2
Defective power supply.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Lamp disconnected.
Ensure the lamp is connected.
2
Defective or non-Agilent lamp.
Exchange the lamp.
3
Defective power supply.
Please contact your Agilent service
representative.
4
Defective main board.
Please contact your Agilent service
representative.
Summary of Contents for G1314B
Page 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Page 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Page 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Page 207: ......