128
Agilent 1260 Infinity VWD User Manual
9
Maintenance
Repairing the Flow Cells
Repairing the Flow Cells
Disassembling the Flow Cell
1
Unscrew the cell screw using a 4- mm hexagonal wrench.
2
Remove the SST rings using a pair of tweezers.
3
Use adhesive tape to remove the peek ring, the window and the gasket.
4
Repeat step a through step c for the other window (keep the parts
separate - otherwise they could be mixed!).
When
If the flow cell needs repair due to leaks or contaminations.
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Wrench, 4 mm hexagonal
Toothpick
Parts required
Description
See
“Standard Flow Cell 10 mm / 14 µL”
See
See
“Micro Flow Cell, 5 mm / 1 µL (only for support)”
See
“Semi-micro Flow Cell 6 mm / 5 µL”
See
“High Pressure Flow Cell 10 mm / 14 µL”
Preparations
•
Turn off the flow.
•
Remove the front cover.
•
Remove the flow cell, see
N O T E
The shown cell parts will differ depending upon the flow cell type. For detailed parts
schematics, refer to above mentioned pages.
C A U T I O N
Scratched window surfaces by tweezers
Window surfaces can easily be scratched by using tweezers for removing the
windows.
➔
Do not use tweezers to remove windows