Agilent 1260 Infinity VWD User Manual
129
Maintenance
9
Repairing the Flow Cells
Cleaning the Flow Cell Parts
1
Pour isopropanol into the cell hole and wipe clean with a piece of
lint- free cloth.
2
Clean the windows with ethanol or methanol. Dry it with a piece of
lint- free cloth.
Reassembling the Flow Cell
1
Hold the flow cell cassette horizontally and place gasket in position.
Ensure both cell holes can be seen through the holes of gasket.
2
Place the window on gasket.
3
Place the peek ring on the window.
4
Insert the conical springs. Make sure the conical springs point towards
the window. Otherwise tightening the cell screw might break the
window.
Figure 41
Orientation of conical springs
5
Screw the cell screw into the flow cell and tighten the screw.
6
Repeat the procedure for the other cell side.
Next steps
1
Reconnect the capillaries.
2
Perform a leak test. If OK, insert the flow cell.
3
Perform
“Wavelength Verification- Calibration”
on page 108 to check the
correct positioning of the flow cell.
4
Replace the front cover.
N O T E
Always use new gaskets.
N O T E
The semi-micro #1 and #2 gaskets (items 6 and 7,
“Semi-micro Flow Cell 6 mm / 5 µL”
page 144) look very similar. Do not mix them up.
8dc^XVaheg^c\h
G^c\"L^cYdl"<Vh`Zi"6ggVc\ZbZci
Summary of Contents for G1314B
Page 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Page 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Page 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Page 207: ......