Agilent 1260 Infinity VWD User Manual
125
Maintenance
9
Exchanging a Lamp
5
Replace the front cover.
Next Steps:
6
Reset the lamp counter as described in the User Interface
documentation.
7
Turn the lamp ON.
8
Give the lamp more than 10 min to warm-up.
9
Perform
“Wavelength Verification-Calibration”
page 108 to check the correct positioning of the lamp.
Summary of Contents for G1314B
Page 1: ...Agilent Technologies Agilent 1260 Infinity Variable Wavelength Detector User Manual...
Page 102: ...102 Agilent 1260 Infinity VWD User Manual 7 Error Information Detector Error Messages...
Page 164: ...164 Agilent 1260 Infinity VWD User Manual 11 Identifying Cables External Contact Cable...
Page 207: ......