54
Agilent 1260 Infinity VWD User Manual
4
Using the Detector
Setting up an Analysis
Conditions
A single injection of the isocratic test standard is made under the
conditions given in
Table 8
Conditions
Flow
1.5 mL/min
Stoptime
8 min
Solvent
100% (30% water/70% Acetonitrile)
Temperature
Ambient
Wavelength
sample 254 nm
Injection Volume
1 µL
Column Temperature (optional):
25 °C or ambient