Power-On Condition: No waveform segments are defined
Example
Deleting a Waveform
LIST:SEL ABC
Selects waveform segment ABC.
LIST:DEL
Deletes segment.
[:SEGMent]:FREE?
[SOURce:]LIST[1][:SEGMent]:FREE?
returns information on waveform segment
memory availability and usage. The return data format is:
<numeric_value>,<numeric_value>
The first numeric value shows the amount of waveform segment memory available
in points; the second, the amount of waveform segment memory used in points.
Comments
•
Executable when Initiated: Yes
•
Coupling Group: None
•
*RST Condition: None
•
Power-On Condition: All of the waveform segment memory is available
Example
Querying Waveform Segment Memory Usage
LIST:FREE?
Queries segment memory usage.
[SOURce:]LIST[1]