[:DC]:POINt?
CALibration[:DC]:POINt? <
value
>
takes the measured value for the current DC
calibration point, computes needed calibration constants, and sets up the
Agilent E1445A for the next measurement. When all measurements have been
made, the calibration constants are checked for validity. If the validity check passes,
the constants are stored in the Agilent E1445A’s non-volatile calibration memory
and the calibration count (
CALibration:COUNt?
query) is incremented.
The
*RST
command should be sent after completing the calibration procedure to
restore normal operation.
Calibration security must have been previously disabled. See the Agilent E1445A
Service Manual for detailed information on the use of this command.
Comments
•
Most of the Agilent E1445A’s commands cannot be executed while calibration is
in progress. The
*RST
command may be used to prematurely terminate the
calibration procedure without affecting the stored calibration constants.
•
Executable when Initiated: No
•
Coupling Group: None
•
Related Commands:
CALibration[:DC]:BEGin
,
CALibration:SECure[:STATe]
•
*RST Condition: None
CALibration
Chapter 8
Command Reference 301
Summary of Contents for E1445A
Page 2: ......
Page 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Page 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Page 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 54: ...Notes 52 Getting Started Chapter 1...
Page 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Page 204: ...Notes 202 Arming and Triggering Chapter 5...
Page 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Page 446: ...Notes 444 AFG Status Chapter 9...
Page 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Page 534: ...Notes 532 Agilent E1445A User s Manual Index...