[:LEVel][:IMMediate][:AMPLitude]:UNIT[:VOLTage]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:UNIT[:VOLTage] <
units
>
sets the default units for subsequent
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
commands.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
units
>
discrete
DBM | DBMW | V | VPK | VPP |
VRMS | W
none
Comments
•
The available default units are:
–
DBM | DBMW:
dB referenced to 1 milliwatt.
–
V:
Volts. This is equivalent to VPK for time-varying waveforms.
–
VPK:
Volts peak
–
VPP:
Volts peak-to-peak
–
VRMS:
Volts RMS
–
W:
Watts
For W, DBM, and DBMW, the amplitude is referenced to the
OUTPut[1]:LOAD
value; they are meaningless and therefore unavailable if
OUTPut[1]:LOAD INFinity
is set.
•
Executable when Initiated: Yes
•
Coupling Group: None
•
Related Commands:
OUTPut[1]:IMPedance
,
OUTPut[1]:LOAD
,
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
•
*RST Condition:
SOURce:VOLTage:LEVel:IMMediate:AMPLitude:UNIT:VOLTage V
Example
Setting the Default Voltage Units
VOLT:UNIT:VOLT VPP
Sets default units to volts peak-to-peak.
[SOURce:]VOLTage
Chapter 8
Command Reference 379
Summary of Contents for E1445A
Page 2: ......
Page 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Page 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Page 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 54: ...Notes 52 Getting Started Chapter 1...
Page 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Page 204: ...Notes 202 Arming and Triggering Chapter 5...
Page 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Page 446: ...Notes 444 AFG Status Chapter 9...
Page 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Page 534: ...Notes 532 Agilent E1445A User s Manual Index...