[SOURce:]VOLTage
The
[SOURce:]VOLTage
subsystem controls the amplitude and offset values for all
output waveform shapes.
Subsystem Syntax
[SOURce:]
VOLTage
[:LEVel]
[:IMMediate]
[:AMPLitude] <
amplitude
>
:UNIT
[:VOLTage] <
units
>
:OFFSet <
offset
>
[:LEVel][:IMMediate][:AMPLitude]
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
amplitude
>
sets the
output amplitude when
SOURce:FUNCtion[:SHAPe]
is set to
DC
,
RAMP
,
SINusoid
,
SQUare
, or
TRIangle
. It sets the positive full-scale output amplitude for arbitrary
waveforms (
SOURce:FUNCtion[:SHAPe] USER
set); the least significant DAC
code bit represents 1/4095 of this value.
Output amplitude for ramp, sine, square and triangle wave output may be programmed
in volts, peak volts, peak-to-peak volts, RMS volts, or dBM. Output amplitude for DC
must be programmed in volts; for arbitrary waveform output, volts or peak volts.
The query form returns the amplitude in terms of the default units, specified by the
SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude]:UNIT[:VOLTage]
command.
[SOURce:]VOLTage
Chapter 8
Command Reference 377
Summary of Contents for E1445A
Page 2: ......
Page 14: ...Notes 12 Agilent E1445A User s Manual Contents...
Page 18: ...Notes 16 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 19: ...Notes Agilent E1445A Arbitrary Function Generator User s Manual 17...
Page 20: ...Notes 18 Agilent E1445A Arbitrary Function Generator User s Manual...
Page 54: ...Notes 52 Getting Started Chapter 1...
Page 84: ...Notes 82 Generating Standard Waveforms Chapter 2...
Page 204: ...Notes 202 Arming and Triggering Chapter 5...
Page 432: ...Figure 9 1 E1445A Status Groups and Associated Registers 430 AFG Status Chapter 9...
Page 446: ...Notes 444 AFG Status Chapter 9...
Page 464: ...Notes 462 Agilent E1445A Specifications Appendix A...
Page 534: ...Notes 532 Agilent E1445A User s Manual Index...