186
Agilent E1439 Programmer's Reference
Functions listed alphabetically
AGE1439_EXTERNAL
uses transitions on the signal applied to the BNC external trigger
connector on the front panel.
AGE1439_EXTERNAL_ECL
uses ECL level transitions on the signal applied to the BNC
external trigger connector on the front panel.
Note
AGE1439_EXTERNAL_ECL
has the same constant value as the
AGE1439_EXTERNAL
constant in the E1439C.
AGE1439_EXTERNAL
is retained for backward compatibility.
AGE1439_EXTERNAL_TTL
uses TTL level transitions on the signal applied to the BNC
external trigger connector on the front panel.
Note
AGE1439_EXTERNAL_TTL
is supported on all E1439B, C and D modules, but it is not
supported on early E1439A modules. A module with a serial number lower than US41140000
will result in the error
AGE1439_TTL_TRIGGER_NOT_SUPPORTED
.
AGE1439_USER
disables the module from any event-driven trigger generation though it is still
possible to force the module to trigger a measurement by pulling the Sync line once the module is
in the trigger state. You may do this by calling the
age1439_meas_start
function, waiting for the
module to reach the trigger state, then triggering the measurement by using
age1439_meas_
control
to pull the Sync line.
AGE1439_IMMEDIATE
triggers a measurement immediately upon entering the trigger state.
Note
In multi-module systems all modules should be use the same trigger type in order to have the
same actual delay.
trigTypePtr
points to the current value of
trigType
.
Return Value
AGE1439_SUCCESS
indicates that a function was successful.
Values other than
AGE1439_SUCCESS
indicate an error condition or other important status
condition. To determine the error message, pass the return value to
“age1439_error_message” on
page 102
.
See Also
“Commands which halt active measurements” on page 198
,
“Default values” on page 201
,
“age1439_init” on page 132
,
“age1439_frequency_setup” on page 128
,
“age1439_data_setup” on
page 90
,
“age1439_data_scale_get” on page 89
,
“age1439_filter_setup” on page 120
,
“age1439_
meas_start” on page 155
,
“age1439_meas_control” on page 151
,
“age1439_trigger_phase_
actual_get” on page 182
,
“age1439_trigger_delay_actual_get” on page 181
,
“The measurement
loop” in chapter 3
,
“Managing multiple modules” in chapter 3
,
“Delay and phase in triggered
measurements” in chapter 3
,
“Magnitude trigger and magdwell time” in chapter 3
Summary of Contents for E1439
Page 6: ...6...
Page 10: ...10 Contents...
Page 11: ...1 1 Installing the Agilent E1439...
Page 18: ...8 Installing the Agilent E1439 To transport the module...
Page 19: ...2 2 Getting Started with the Agilent E1439...
Page 28: ...18 Getting Started with the Agilent E1439 To use the example programs...
Page 29: ...3 3 Using the Agilent E1439...
Page 63: ...4 4 Agilent E1439 Programmer s Reference...
Page 217: ...5 5 Module Description...
Page 228: ...218 Module Description Block diagram and description...
Page 229: ...6 6 Replacing Assemblies...
Page 234: ...224 Replacing Assemblies Replaceable parts To remove the top cover...
Page 235: ...225 Replacing Assemblies Replaceable parts To remove the M1 M2 assemblies...
Page 236: ...226 Replacing Assemblies Replaceable parts...
Page 238: ...228...
Page 244: ...Index 234...