![Agilent Technologies E1439 User Manual Download Page 101](http://html1.mh-extra.com/html/agilent-technologies/e1439/e1439_user-manual_2868333101.webp)
91
Agilent E1439 Programmer's Reference
Functions listed alphabetically
AGE1439_BLOCKSIZE_MAX
selects the maximum blocksize.
AGE1439_BLOCKSIZE_DEF
sets the default blocksize.
The range of available block sizes depends on the number of bytes required for each sample. The
command accepts any number between 2 and memory size (in bytes)
×
2/3. If the requested block
size falls outside the range shown in the table the previous valid value is used and a status register
flag (bit 6) is set indicating a setup error. The blocksize is updated after the setup is changed to be
valid.
For real data
blocksize
is the number of real data values per data block. For complex data
blocksize
is the number of complex data pairs per data block.
The following table summarizes the available block sizes for each setting of the
dataType
, and
resolution
parameters.
Note
Block size must be an even number. Considerably more samples may need to be taken in order to
set the block available status bit.
blocksizePtr
points to the current value of the
blocksize
parameter. The returned value is the closest valid value
to the requested block size.
dataDelay
is used to specify the minimum FIFO delay in number of samples. This parameter applies only in
continuous mode.
AGE1439_DATA_DELAY_MAX
sets the maximum allowable delay.
AGE1439_DATA_DELAY_MIN
sets the minimum allowable delay.
dataDelayPtr
points to the current value of the
delay
parameter.
dataType
determines whether the Agilent E1439 collects and returns real or complex data.
Setting this parameter to
AGE1439_REAL
causes only the real part of the data to be returned for
each sample
AGE1439_COMPLEX
causes the real data followed by the imaginary data to be returned in each
sample.
Normally, if the center frequency set with the
age1439_frequency_setup
function is zero, the
type should be set to
AGE1439_REAL
since the imaginary component of each sample is zero
anyway. When non-zero center frequencies are used the type should normally be set to
AGE1439_COMPLEX
, otherwise the imaginary component of the signal is lost.
data type
resolution
min. block size
max block size
in Msamples
(2 M*72 memory)
1
1. Parity memory is used in non-parity mode, so 2M
×
72
bit memory yields 18 Mbytes of FIFO storage.
real
12
6
12
real
24
3
6
complex
12
3
6
complex
24
2
3
Summary of Contents for E1439
Page 6: ...6...
Page 10: ...10 Contents...
Page 11: ...1 1 Installing the Agilent E1439...
Page 18: ...8 Installing the Agilent E1439 To transport the module...
Page 19: ...2 2 Getting Started with the Agilent E1439...
Page 28: ...18 Getting Started with the Agilent E1439 To use the example programs...
Page 29: ...3 3 Using the Agilent E1439...
Page 63: ...4 4 Agilent E1439 Programmer s Reference...
Page 217: ...5 5 Module Description...
Page 228: ...218 Module Description Block diagram and description...
Page 229: ...6 6 Replacing Assemblies...
Page 234: ...224 Replacing Assemblies Replaceable parts To remove the top cover...
Page 235: ...225 Replacing Assemblies Replaceable parts To remove the M1 M2 assemblies...
Page 236: ...226 Replacing Assemblies Replaceable parts...
Page 238: ...228...