Required Status
Groups
All SCPI instruments must implement a minimum set of status groups.
Some instruments contain additional status groups, consistent with the
general status register model. The minimum required status system is
shown in Figure 4-5.
The Standard Operation Status and Questionable Data Groups are 16-bits
wide, while Status Byte and Standard Event Groups are only 8-bits wide. In
all 16-bit groups, the most significant bit (bit 15) is not used. Bit 15 always
returns a zero. The commands that set and query bits in the Status Registers
all use decimal integers. For example, you send
*ESE 4
to set bit 2 of the
Standard Event Enable Register. Similarly, a response of "8" to the query
*ESE?
indicates that bit 3 is set. The remainder of this chapter explains
each status group in detail.
Figure 4-5. Minimum Required Status Register System
106 Triggering and System Status
Chapter 4
Summary of Contents for E1406A
Page 2: ......
Page 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Page 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Page 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Page 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Page 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Page 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...