:CONFigure:MTABle
VXI:CONFigure:MTABle <
address
>
links a user-defined A24/A32 address
allocation table to the command module (resource manager) processor. The
command module must be the acting resource manager in order for the table
to be implemented.
Parameters
Parameter
Name
Parameter
Type
Range of Values
Default
Units
<
address
>
numeric
(DIAG:NRAM:ADDR?)
none
Comments
•
Be certain that address specifies the starting address of the area in
user RAM (allocated using
DIAG:NRAM:CREate
) where you stored
the A24/A32 address allocation table data.
•
Tables must start on an even address. Note that
DIAG:NRAM:CREate
allocates RAM for the table with an even starting address.
•
<
address
>
may be specified in decimal, hex (
#H
), octal (
#Q
), or
binary (
#B
) formats.
•
Setting address to 0 prevents the parameters defined by the table
from being invoked when the system is rebooted, however, the table
remains in user RAM.
•
For more information see “ Reserving A24/A32 Address Space” on
page 48.
•
Related Commands:
DIAG:NRAM:CREate
,
DIAG:NRAM:ADDRress?
,
DIAG:DOWNload
,
VXI:CONF:MTABle?
Example
Link an A24/A32 Address Allocation Table to the Processor
DIAG:NRAM:CRE <
size
>
Allocate space for table in user RAM.
DIAG:BOOT
Reboot system to complete the
allocation.
DIAG:NRAM:ADDR?
Get starting address of table
(RAM segment).
DIAG:DOWN
<
address
>
,
<
data
>
Download data into table.
VXI:CONF:MTAB
<
address
>
Link table to processor.
DIAG:BOOT
Reboot system to implement table.
Chapter 5
Agilent E1406A Command Reference 201
Summary of Contents for E1406A
Page 2: ......
Page 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Page 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Page 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Page 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Page 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Page 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...