:CONFigure:DLISt?
VXI:CONFigure:DLISt? [<
logical_addr
>]
returns information about
the device specified by logical_addr. Response data is in the form:
n1, n2, n3, n4, n5, n6, c1, c2, c3, c4, c5, s1, s2, s3, s4
Where the fields above are defined as:
n
fields
Indicate numeric data response fields.
c
fields
Indicate character data response fields.
s
fields
Indicate string data response fields.
n1
Device’s Logical Address.
A number from 0 to 255.
n2
Commander’s Logical Address. A number from -1 to 255;
-1 means this device has no commander.
n3
Manufacturer’s ID. A number from 0 to 4095.
n4
Model Code. A number from 0 to 65535, chosen by the manufacturer to
signify the model of this device.
n5
Slot Number. A number between -1 and the number of slots in this
mainframe; -1 indicates that the slot associated with this device is unknown.
This is always -1 for B size mainframes.
n6
Slot 0 Logical Address. A number from 0 to 255.
c1
Device Class. 3 data characters;
EXT|HYB|MEM|MSG|REG|VME
.
EXT
= Extended device,
HYB
= Hybrid device (e.g., IBASIC
MEM
= Memory device,
MSG
= Message-based device
REG
= Register-based device,
VME
= VME device
c2
Memory Space. Up to 4 data characters;
A16|A24|A32|NONE|RES
.
A16
= A16 addressing mode,
A24
= A24 addressing mode,
A32
= A32 addressing mode,
NONE
= no addressing mode,
RES
= reserved.
c3
Memory Offset. 10 data characters which define the base address of the
A24 or A32 address space on the device. This value is expressed in hex
format (first two characters are
#H
).
c4
Memory Size. 10 data characters which define the size of the A24 or
A32 address space in bytes. This value is expressed in hex format
(first two characters are
#H
).
c5
Pass/Failed. Up to 5 data characters which define the status of the device;
FAIL | IFAIL | PASS | READY. FAIL
= failed self-test,
IFAIL
= configuration register initialization fails,
PASS
= self-test passed,
READY
= ready to receive commands
s1
Extended Field 1. Not currently used; returns ""
s2
Extended Field 2. Not currently used; returns ""
s3
Extended Field 3. Not currently used; returns ""
s4
Manufacturer’s Specific Comments. Up to 80 character string contains
manufacturer specific data in string response data format. This field is sent
with a 488.2 string response data format, and will contain the instrument
name and its IEEE 488.1 secondary address unless a start-up error is
detected. In that case, this field will contain one or more error codes in the
form "
CNFG ERROR: n, m, ...,z
". Table B-3 in Appendix B for a complete
list of these codes.
190 Agilent E1406A Command Reference
Chapter 5
Summary of Contents for E1406A
Page 2: ......
Page 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Page 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Page 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Page 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Page 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Page 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...