:WSProtocol
:COMMand:
command
VXI:WSProtocol:COMMand
:command is a series of commands which sends
the specified Word Serial Command to the address set using the
VXI:SELect
command and continues without waiting for a response. The response to this
command can be read with the
VXI:WSProtocol:RESPonse?
command. The
following table lists the available commands and their parameters (if any).
:command
parameter1
parameter2
Description
:AHLine
<
hand_id
>
(1-7)
<
line_#
>
(0-7)
Assign Handler Line. A line number of 0
means the handler is to be disconnected.
:AILine
<
int_id
>
(1 - 7)
<
line_#
>
(0 - 7)
Assign Interrupter Line. A line number of 0
means the handler is to be disconnected.
:AMControl
<
rspns_mask
>
(0 - 15)
Asynchronous Mode Control
:ANO
Abort Normal Operation
:ANY
<
cmd_word
>
(-32768 - 32767)
Specify any word serial command as a
16-bit value in
cmd_word
.
:BAVailable
<
end_bit
>
(1 | 0 | OFF | ON)
<
byte
>
(0 - 255)
Byte Available
(bit 8 = 1 = END, bits 7 - 0 = data byte)
:BNO
<
top_level
>
(1 | 0 | OFF | ON)
Begin Normal Operation
:BREQuest
Byte Request
:CEVent
<
enable
>
(0 | 1 | OFF | ON)
<
event_#
>
(0 - 127) Control Event
:CLEar
Clear
:CLOCk
Clear Lock
:CRESponse
<
rspns_mask
>
(0 - 127)
Control Response
:ENO
End Normal Operation
:GDEVice
<
cmdr_laddr
>
(0 - 255)
Grant Device
:ICOMmander
Identify Commander
:RDEVice
<
logical_addr
>
(0 - 255)
Release Device
:RHANdlers
Read Handlers
:RHLine
<
hand_id
>
(1 - 7)
Read Handler Line
:RILine
<
int_id
>
(1 - 7)
Read Interrupter Line
:RINTerrupter
Read Interrupters
:RMODid
Read MODID
:RPERror
Read Protocol Error
:RPRotocol
Read Protocol
:RSARea
Read Servant Area
:RSTB
Read STB
:SLModid
<
enable
>
(0 | 1 | OFF | ON)
<
modid
>
(0 - 127)
Set Lower MODID (lines 0 - 6)
:SLOCk
Set Lock
:SUModid
<
enable
>
(0 | 1 | OFF | ON)
<
modid
>
(0 - 63)
Set Upper MODID (lines 7 - 12)
:TRIGger
Trigger
214 Agilent E1406A Command Reference
Chapter 5
Summary of Contents for E1406A
Page 2: ......
Page 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Page 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Page 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Page 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Page 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Page 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...