Reserving A24/A32
Address Space
As previously mentioned, the resource manager cannot determine when
VME devices have been installed in the system. To prevent the resource
manager from allocating A24/A32 addresses intended for VME devices to
VXIbus devices, the address allocation table is used. The A24/A32 address
allocation table is also used to assign different addresses to VXIbus devices
other than those (default) addresses assigned by the resource manager
during power-on.
The A24/A32 Address
Allocation Table
The A24/A32 address allocation table is created and stored in the command
module as follows:
1. Table space in the command module’s non-volatile user RAM is
made available by allocating a segment of RAM with the command:
DIAGnostic:NRAM:CREate <
size
>
2. Reset the command module.
NRAM
is created during the boot-up
process:
DIAGnostic:BOOT:WARM
3. The location (starting address) of the table in RAM is determined
with the command:
DIAGnostic:NRAM:ADDRess?
4. Data is downloaded into the table with the command:
DIAGnostic:DOWNload <
address
>, <
data
>
5. The table is linked to the appropriate algorithm in the command
module processor with the command:
VXI:CONFigure:MTABle <
address
>
Table Format
The format of the A24/A32 address allocation table is shown in Table 2-6.
Table 2-6. A24/A32 Address Allocation Table Format
Table Format
Memory Record
Format
Valid Flag/
Number of Records
Laddr
Address Record #1
Frame ID Addr space
Address Record #2
Base addr
•
Memory size
•
Address Record N
48 Configuring the Agilent E1406A Command Module
Chapter 2
Summary of Contents for E1406A
Page 2: ......
Page 14: ...Notes 12 Agilent E1406A Command Module User s Manual ...
Page 15: ...Notes Agilent E1406A Command Module User s Manual 13 ...
Page 16: ...Notes 14 Agilent E1406A Command Module User s Manual ...
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 120: ...Notes 118 Triggering and System Status Chapter 4 ...
Page 179: ... DIAG BOOT COLD will set PARity to NONE Chapter 5 Agilent E1406A Command Reference 177 ...
Page 240: ...Notes 238 Agilent E1406A Command Reference Chapter 5 ...
Page 252: ...Notes 248 Agilent E1406A Specifications and General Information Appendix A ...